Nanoporous materials as carriers of hydrogen peroxide vapour: A new bio-decontamination technology


Fadhil Musa¹, Raquel Nogueira¹, Margarida Beiral¹, Fernando Antunes², João Pires²* | EJPPS | 31106 | DOI:https://doi.org/10.37521/ejpps31112



Abstract

High-level bio-decontamination, which involves reducing microorganisms by 99.9999%, is essential in preventing Hospital Acquired Infections and controlling pandemics. This study demonstrates that nanoporous materials, which can retain molecules within their pores, and subsequently release them, can be used in high level bio-decontamination. H₂O₂ vapour is a golden-standard in high level bio-decontamination. This work demonstrates that various nanoporous materials, particularly mesoporous silicas, can be utilized to store and release H₂O₂ in the vapour phase. H₂O₂ concentrations of over 2500ppm were achieved by desorbing it from the carrier material at low temperatures of 60–80°C. Generation of H₂O₂-vapour by desorption from nanoporous materials is technically much simpler than vaporization of aqueous H₂O₂ solutions, which use flash vaporization processes occurring at 130–50°C. This has important technical implications, highlighting the potential of nanoporous materials as carriers for H₂O₂ for high-level bio-decontamination.


Keywords: Hydrogen peroxide, Vaporization, Bio-decontamination, Nanoporous materials, Sterilization, Silica-gel


Introduction

Long before the recent COVID-19 pandemics, there was an increasing awareness that improved cleaning and disinfection of environmental surfaces is needed in healthcare facilities, as previously reviewed in the literature¹⁻³. An important example is the hospital-acquired infections (HAI) since various multi resistant pathogens can persist in the healthcare environment for days⁴. Only in American hospitals, the Centre for Disease Control estimates that HAI account for an estimated

1.7 million infections and 99,000 associated deaths per year. Bio-decontamination is relevant in other important sectors as aseptic packaging processes⁵ or the maintenance of biological safety cabinets⁶.

Traditional cleaning methods (detergent- and disinfectant-based) are notoriously inefficient for decontamination and new approaches have been proposed in recent years. Among these, the more relevant ones are devices that use ultraviolet (UV) light, formaldehyde, chlorine dioxide, peracetic acid, ozone, and hydrogen peroxide (H₂O₂) vapour. All these methodologies have their advantages and disadvantages. For instance, UV light based apparatus allow short bio-decontamination cycles of bio-decontamination, but are expensive and, more importantly, shadow areas are not decontaminated⁴. Formaldehyde and peracetic acid are effective, but they are also highly toxic, carcinogenic, and leave residues. Chlorine dioxide is effective and non-carcinogenic, but some limitations on venting to the atmosphere may apply⁶. H₂O₂ vapour is highly effective, non-carcinogenic, and does not leave residues, since it decomposes into water and oxygen²,⁶.

A drawback in the use of H₂O₂ vapour is due to technological issues related with its vaporization that render the final bio-decontamination equipment expensive and bulky. In fact, present technologies use aqueous solutions, where the H₂O₂ vapour that can be delivered is obviously limited by the Raoult’s law and the ratio of H₂O/H₂O₂ vapour is even higher in the vapour than in the solution⁷. To improve peroxide concentration in the vapour phase, and at least ensure that the H₂O/H₂O₂ ratio in the vapour is similar to the original liquid solution, flash vaporization is employed⁸.

This has two major consequences:

(i) it makes the final equipment expensive, limiting the general use of a bio-decontamination procedure that has proved a high efficacy in reducing 10⁶ times (6-Log reduction or 99.9999% reduction) even the more resistant pathogens, including Mycobacterium tuberculosis, Mycoplasma, Acinetobacter, C. difficile, Bacillus anthracis, viruses, prions and fungi (ex. Candida auris)³,⁹.

(ii) the equipment lacks compactness, which precludes the use of the method for applications in small spaces.

Nanoporous materials possess porous with diameter between 1 and 100 nm¹⁰. According to IUPAC, nanoporous materials can be classified by pore size, in 3 categories: microporous materials (0–2-nm pores), mesoporous materials (2–50-nm pores), and macroporous materials (>50-nm pores)¹⁰. These materials have many applications such as adsorbents catalysts and ion exchangers¹⁰⁻¹². In this work, we present an innovative methodology that utilizes inexpensive nanoporous materials to store and release H₂O₂ vapour for use in bio-decontamination. To the best of our knowledge this methodology has not been previously described in the literature. It will be shown that simple, affordable, and compact devices for bio-decontamination with H₂O₂ can be obtained without any loss of efficiency. These devices can operate in large spaces but are especially adequate for uses in small spaces that range from gloveboxes, bio-safety cabinets, fume hoods, chambers with personal equipment for domestic applications (as masks or cell phones) and terminal sterilization of solid biopharmaceutical molecules.


2. Experimental


2.1. Materials

A screening of various types of nanoporous materials was made, for their capacity to store and release H₂O₂. A partial list of the tested materials, selected based on their general availability and performance, is shown in Table 1. In Supplementary Information (Section 1)a more complete list of the studied materials, which included clay-based materials and carbon materials, is presented.

To evaluate the specific surface area and porous volumes the usual method of nitrogen adsorption at - 196°C was made for selected samples, using an automated apparatus (ASAP 2010 from Micromeritics or NOVA 2200 from Quantachrome). Prior to the adsorption measurements, the samples were outgassed (under vacuum) at 300°C for 2.5 h.


2.2. Loading of the materials with H₂O₂


2.2.1. Loading of H₂O₂ from liquid solution

Porous materials loading was initially made by mixing the powder with a 30% H₂O₂ solution (Acros Organic, 35%) in a 1:1 vol ratio for a given time. To improve the separation of the solid material from the

liquid H₂O₂ solution, an attempt was made by separating the solid from the liquid, using a dialysis tubing cellulose membrane (Sigma-Aldrich, avg. flat width of 43 mm).


2.2.2. Loading of H₂O₂ from vapour – static mode

The materials, previously weighted and dried in an oven at 100°C, dispersed in the base of small flasks, were then placed in a desiccator saturated with H₂O₂ vapour. This saturation was achieved by the evaporation of a 35% H₂O₂ solution placed at the bottom of the desiccator. The desiccator was covered with aluminium foil to avoid H₂O₂ photodecomposition. For larger amounts of powder, a closed polypropylene box was used. Loading times between 1 and 4 days were tested. The experiments were made at ambient temperature.


Table 1

Materials used for H₂O₂ storage and release tests (a more complete list is

given in Supplementary Information).

Material

Observations

Mesoporous Silicas


SBA-15


Synthesised as described in¹³

MCM-41

Synthesised as described in¹⁴

SG60

Silica gel from ACROS ORGANICS

Zeolites


3A

Potassium A zeolite from BDH

4A

Sodium A zeolite from Sigma-Aldrich

NaX

13X zeolite from BDH

NaY

Y zeolite from Aldrich

NH4X

Prepared from 13X by cation exchange


2.2.3. Loading of H₂O₂ from vapour – continuous mode

To prepare large amounts of porous materials, loaded with H₂O₂, a continuous method was developed. Basically, the approach involved a container only partially filled with liquid H₂O₂, to allow evaporation of H₂O₂ and its removal by a continuous air flow. The air flow (1.2 m³/h) was achieved with a vacuum pump (Diaphragm Vacuum Pump Laboport, 100% oil free, Politetrafluoretileno coated) and made to pass in a glass container (with a volume of 450 cm³), where the powder to be loaded (50g) was placed and retained with a welded porous glass filter.


2.3. Determination of the H₂O₂ content

H₂O₂ content after loading in a given solid material was done by H₂O₂ solubilization in water and measuring through an electrochemical assay or spectroscopic assay.


2.3.1. Hydrogen peroxide electrode

For determination of the H₂O₂ content in some samples from the static loading, a hydrogen peroxide electrode (WPI, ISO-HPO-100L) was used. Each solid sample was put in 80mL of water and the electrode monitored the change of signal corresponding to the release of hydrogen peroxide. H₂O₂ standard solutions were used to build five-point linear calibration curves required for calculations.


2.3.2. Oxygen electrode

For determination of the H₂O₂ content in samples from continuous loading and stability studies a dissolved oxygen electrode (Oxygraph, Hansatech Instruments) was used. The solid material was put in water and the mixture was filtered (0.45μm Nylon filters) and diluted (10⁻³) before measurement. H₂O₂ standard solutions were used to build a fivepoint linear calibration curves required for calculations. For both, samples and standards, 1mL was added to the chamber, followed by 20μL of a 1mg/L catalase (catalase from bovine liver, ≥10 000 units/mg protein, Sigma Aldrich) solution. The increase in signal after catalase addition, due to the decomposition of H₂O₂ to oxygen and water, was recorded via Oxygraph™ software (Hansatech Instruments) and considered for calculation.


2.3.3. Absorbance

For determination of the H₂O₂ content in some samples from the static loading, the absorbance at 240 nm of the samples was measured. Each solid sample was put in water and the mixture was filtered (0.45μm Nylon filters) and diluted (10⁻³) before measurement. A Jenway7205 UV–Visible spectrophotometer was used.


2.4. Release of H₂O₂ from the materials

For the release of H₂O₂ from the adsorbent materials, a dedicated equipment was constructed – Fig. S1 in Supplementary Information. Briefly, this equipment (a heating unit) has a heating module with a fan that produces an air flow with a controlled temperature (temperature values between 60 and 80°C were used). The air flow passes through a cartridge, that contains the powder loaded with H₂O₂, desorbing it, and carrying the H₂O₂ in the vapour state to the surroundings. Efficiency, as discussed below, was measured by the bio-decontamination of a biosafety cabinet and a glove box. The measurement of the H₂O₂ concentration in the vapour phase, as well as the measurement of humidity, was made through a dedicated H₂O₂ probe from VAISALA, model PEROXCAP® Probe HPP272.


2.5. Evaluation of efficacy

2.5.1. Chemical indicators

H₂O₂ vapour chemical indicators (type 4, gke Steri-Record, gke-GmbH, Germany) were used in various situations (ex.various locations inside a glove box where the bio-decontamination cycles were carried

out – see below) allowing for the evaluation of the process of releasing the hydrogen vapour from the adsorbent materials.


2.5.2. Biological indicators

To evaluate the efficacy of microbial sterilization, Geobacillus stearothermophilus spore discs containing 10⁶ spores inoculated in a glass fibre carrier were purchased from gke-GmbH, Germany. At the end of the bio-decontamination cycle, the biological indicators were aseptically transferred into Tryptic Soy Broth (TSB) with pH-indicator (gke Steri-Record, gke-GmbH, Germany) and incubated at 60°C, for 7 days. Change of medium colour indicated bacterial growth and failure of the decontamination process.

For the determination of D-value, number of colony units remaining after the decontamination was done with the Most Probable Number method¹⁵,¹⁶. The spores were put in reverse osmosis (RO) water and the samples were vortexed and treated in an ultrasonic bath for 30 min, to remove the spores from the carrier and singularize them in the suspension. The resulting solutions were serial diluted in RO water to a level of 10⁻⁶. 10μL of each dilution was inoculated with 190μL of growth medium in each well of the plate and incubated at 60°C. The resulting data were interpreted by the means of Most Probable Number tables¹⁷. Three conditions were evaluated using three peak H₂O₂ concentrations: 379 parts per million (ppm) – 5g of SG-60 material, glovebox temperature was 24.5–30.0°C; 468ppm–5g of SG 60 material, glovebox temperature was 35°C throughout the experiment; 2578ppm–20g of SG-60 material, glovebox temperature was between 40 and 60°C throughout the experiment.


2.6. Stability of the stored H₂O₂

To improve the shelf life of the adsorbed H₂O₂, that is to decrease H₂O₂ loss due to desorption and/or decomposition over time, stabilizers were incorporated in SG60. Various types of stabilizers were used such as: sodium phosphate; sulfuric acid; urea, sodium stannate, sodium oxalate, Ethylenediaminetetraacetic acid (EDTA, ThermoFisher, 99.5%), Citric acid (≥99%, Sigma Aldrich), Diethylenetriaminepentaacetic acid (DTPA, ≥98%, Sigma Aldrich). The general approach was to physically mix the stabilizer with the adsorbent materials in various proportions, from 1 to 20% (in weight), before H₂O₂ vapour loading. Stability of the final product was evaluated by determining the H₂O₂ retained by the material after a given period in three different storing conditions: ambient temperature, fridge (8°C) and freezer (⁻18°C).


2.7. Evaluation of reusability of silica

To determine the reusability of silica SG60, specifically whether it could be recharged multiple times, after being used, i.e., after the loading and release of H₂O₂, the silica was subjected to a drying and reloading process. Loading in continuous mode was used.


2.8. DNAse bio-decontamination and activity determination

The bio-decontamination procedure involved exposing an open vial containing approximately 2000 Kunitz of Deoxyribonuclease (DNAse, D4263, Sigma-Aldrich, Deoxyribonuclease I from bovine pancreas) to 100–300mg of H₂O₂-loaded material in a 50mL container for seven days at ⁻ 18.4°C. Inside the container was also a biological indicator, which was aseptically incubated in TSB media after the 7 days. To evaluate the effect of the exposure of DNase the H₂O₂-loaded material DNAse activity was determined by following absorbance at 260nm for 5 min in a reaction mix containing 83mM Sodium Acetate (S8625, Sigma-Aldrich), 4.2 mM Magnesium Sulphate (M1880, Sigma- Aldrich), 0.14% Sodium Chloride (S0817, Sigma-Aldrich), 0.003% (w/ v) Deoxyribonucleic Acid (D3664, Sigma-Aldrich) and 250 Kunitz of DNAse.

For heat inactivation, as a negative control, a solution of DNAse (500 Kunitz/mL) was subjected to 70°C for 15 min.


3. Results and discussion


3.1. Loading of the materials with H₂O₂


3.1.1. Loading from liquid peroxide solution

The first attempts of loading the porous materials, were made by mixing the powder with the H₂O₂ solution. However, since the materials were fine powders, the separation from the liquid, by filtration or centrifugation, showed losses of material and losses of H₂O₂ loaded. A second attempt made by separating the solid from the liquid using a dialysis membrane did not solve this difficulty. In fact, because the water could permeate the dialysis membrane, in the end the material was wet.

The process of drying the material (for instance, in an oven) implied dramatic losses in the loaded H₂O₂. Additionally, the loading yields from a liquid hydrogen peroxide solution were always very low, about 0.03% in total weight for the best cases. These low loading yields were also reported in the literature by other authors¹⁸, which also employ aqueous H₂O₂ solutions as a source. In this way, the loading from liquid hydrogen peroxide solutions was not pursued further.


3.1.2. Loading of H₂O₂ from vapour – static mode

The amounts of H₂O₂ loaded from vapour phase (static mode for three days) in the mesoporous silicas and zeolite 4A are presented in Table 2, in percent of the total mass. The amounts were measured by the methodology involving the H₂O₂ electrode or through absorbance as described in section 2.3.1. Three repetitions were made for each assay and a typical result is exemplified in Fig. 1. Additional quantitative assays were conducted using the absorbance method as outlined in section 2.3.3.

As registered in Table 2, the material with highest loading is themesoporous silica SBA-15, closely followed by MCM-41 and SG60. Zeolite 4A demonstrates a notably reduced H₂O₂ loading capacity, which can likely be attributed to its pronounced hygroscopic nature. Various surface modifications of SBA-15, to enhance the interactions between the material and the H₂O₂ were tried, as summarized in Supplementary Information – Table S2, but did not improve the loading yields. Additionally, as indicated also in Table S2, other types of materials, such as activated carbons or clay-based materials, were also considered but did not present loading yields higher than the samples presented in Table 2.


Table 2

H₂O₂ loadings at ambient temperature (vapour – static method) for the indicated materials.

Material

Loaded H₂O₂ (%)

SBA-15

20.1 ± 0.3

MCM-41

14.1 ± 0.4

4A

4.7 ± 0.3

SG60

14.9 ± 1.3


3.1.3. Loading of H₂O₂ from vapour – continuous mode

Various types of configurations were tested, as schematically described in Fig. S3 – Supplementary Information. Mode E in Fig. S3 was selected, based on a trade-off of conditions such as the process length (the necessary time required to reach at least 15 % of loading yield) and preventing vapour condensation. Loading with continuous mode incorporated higher amounts of H₂O₂ – 18.2 ± 1.5% – in SG60 after 4 days.


Fig. 1. Detection of the amount of H₂O₂ by the selective electrode (examples for SBA-15 and 4A zeolite).
Fig. 1. Detection of the amount of H₂O₂ by the selective electrode (examples for SBA-15 and 4A zeolite).

3.2. Release of H₂O₂ from the materials

As stated above, for the methodology of storing and releasing H₂O₂ with nanoporous materials to be translated into an affordable technology, it would need to operate at a relatively low temperature. Preferably, at a much lower temperature than the ones at which flash-vaporization operates which can achieve 200°C, a temperature that, in fact, can inclusively favour the decomposition of the H₂O₂ molecules⁷,⁸. Nevertheless, very low releasing temperatures may result in inadequate control of H₂O₂ release, and cause part of the stored vapour to desorb prematurely.

To evaluate the temperature needed to release the H₂O₂ from the materials, thermogravimetric (TG) analyses were made. Fig. 2 a) and b) show the TG, and the respective differential curves (DTG), for a mesoporous material (SBA-15), a microporous material (4A zeolite) and the silica SG60.

As could be expected due to the presence of the extra-framework cations in the zeolite¹⁹, which can promote the interactions with the polar H₂O₂ molecule, the temperature at which most of the H₂O₂ is released, corresponding to the inflexion in the TG signal and the minimum in the DTG curve, is higher for the 4A zeolite than for the SBA-15 material, namely 50 and 164°C, respectively. While the latter temperature is too high, the former is too low. In fact, even if the minimum in the DTG curve of SBA-15 is at 50°C, it can be seen from the shapes of this curve (Fig. 2a) that the fall in the mass loss is very steep. To obviate this, the SG60 sample was considered. The minimum in the DTG curve for SG60 is 87°C (Fig. 2 b) and the decrease in the mass loss is less steep than what was found for the SBA-15 material which, potentially, will allow a better control in the release of the H₂O₂. Additionally, both materials are mesoporous materials, as indicated by their type IV adsorption isotherms of nitrogen at -196°C¹⁰ (Supplementary Information Fig. S4). Even if the mesoporous volume is lower for SG60 (0.79cm³g-1) than for SBA-15 (1.21cm³g-1), both materials present mesopore size distributions with similar maxima of pore sizes (Fig. 3).

To the reasons stated above, and even considering that the capacity of SG60 to store H₂O₂ is lower than the observed for SBA-15, an additional relevant reason favours the selection of the SG60 silica. The biodecontamination devices obtained based on the storage and release of H₂O₂ in porous materials as described here will require significant quantities of material, for instance for room decontamination. Therefore, it is important that the system is based in an affordable material and, presently, 1 kg of SBA 15 costs nearly 150 times more than the same amount of the SG60 silica studied here. An additional feature of SG60 is that this material offers an additional advantage over zeolites in that it possesses an amorphous nature. This characteristic is less toxic, making amorphous silicas safer to handle than the crystalline structure of zeolites²¹.

Fig. 2. a) Thermogravimetric and b) Differential Thermogravimetric curves for the indicated materials loaded with H₂O₂.
Fig. 2. a) Thermogravimetric and b) Differential Thermogravimetric curves for the indicated materials loaded with H₂O₂.
Fig. 3. Mesopore size distributions for the SBA-15 and SG60 samples obtained from the nitrogen adsorption isotherms at -196°C by the Broekhoff-de Boer
Fig. 3. Mesopore size distributions for the SBA-15 and SG60 samples obtained from the nitrogen adsorption isotherms at -196°C by the Broekhoff-de Boer

3.3. Stability of the stored H₂O₂ in the SG60 material

The shelf stability of H₂O loaded in SG60 was evaluated since this is an important characteristic for applications in bio-decontamination, namely for product transportation and storage. Three storage conditions were selected (always in the dark): at room temperature, in a fridge (8°C) and in the freezer (-18°C). Illustrative results, obtained from at least 3 replicated experiments, are given in Fig. 4.

It can be noticed from Fig. 4 that storage at room temperature is clearly unappropriated since, even after one week, only 58% of the initial H₂O is present in the material, which decreases for less than 20% of the initial values after one month. Storing the loaded material in the fridge and, particularly, in the freezer improves the stability.

The decomposition of H₂O₂ is a complex process that can proceed by a number of complex reaction pathways²²,²³. To reduce the H₂O₂ decomposition, mostly for aqueous H₂O₂ solutions, a number of stabilizers had been proposed in the literature between organic and inorganic substances²⁴⁻²⁷. Since adsorbed H₂O₂ is a new field of studies, there were not, to our knowledge, specific stabilizers reported in the literature for this type of system. As so, a trial-and-error study was made with the more common stabilizers used in liquid phase. The general approach, as mentioned in the Experimental Section, was to mix the stabilizer in various proportions with the SG60 sample, using incipient wetness, and evaluating the H₂O₂ retained by the material after a given period of time, in different storing conditions. Since stability in the freezer (-18°C) was achieved for the material even without stabilizers, for a period of six months, the study focused on the improvement of stability at room temperature and at the fridge temperature (8°C).

As mentioned, a large number of stabilizers were tested, and the obtained results are given in the Supplementary Information Section 5. In Fig. 5A, we present the results for the systems that showed the more favourable improvement in the stabilization of H₂O₂ loaded in the SG60 material, for storage at room temperature. These systems were based in the addition of EDTA, DTPA or citric acid. It can be seen from this figure that, although some loss of H₂O₂ is always evident, for a two-month storage period EDTA or citric acid (2% weight) present similar results. However, when the results for storage in the fridge (8°C) are also considered – Fig. 5B, citric acid, at 1 or 2% performs better. The H₂O₂ content of SG60 material mixed with EDTA or citric acid, after storing at 3 months at room temperature, is still sufficient to make the efficient bio-decontamination of small spaces such as biosafety cabinets – as discussed in the following section.

Fig. 4. Percentage of H₂O₂ recovered from loaded SG60 material after one week, one month, or six months stored at room temperature, in the fridge (8 ◦C)
Fig. 4. Percentage of H₂O₂ recovered from loaded SG60 material after one week, one month, or six months stored at room temperature, in the fridge (8 ◦C)

Fig. 5. Percentage of H₂O₂ recovered from SG60 material mix with EDTA, DTPA or Citric Acid, in the indicated weight proportion. Samples were stored at (a) room temperature or (b) in the fridge (8 ◦C).
Fig. 5. Percentage of H₂O₂ recovered from SG60 material mix with EDTA, DTPA or Citric Acid, in the indicated weight proportion. Samples were stored at (a) room temperature or (b) in the fridge (8 ◦C).

3.4. Bio-decontamination efficacy of H₂O loaded in SG60

Next, we characterized the bio-decontamination efficacy of HO₂ desorbed form the SG60 material. The release of H₂O₂ from the porous material to the surroundings to be bio-decontaminated was made using a dedicated unit to heat the powder with hot air flow and desorb the H₂O₂ – Supplementary Information, Section 1. A given amount of powder is enclosed in a cartridge in which the top and bottom are two layers of High Efficiency Particulate Absorbing (HEPA) filter. These layers of HEPA filter retain the powder particles but allow a stream of hot air (between 60 and 80°C) to pass and, therefore, enable the release of H₂O₂ in the vapour phase through the top HEPA filter and into the closed environment to be bio-decontaminated. We started by performing a qualitative assay in a glovebox (total volume of 0.25m³, see Fig. S6) by using a standard procedure that uses chemical and biological (Geobacillus stearothermophilus spore discs) indicators²⁸, as described in the experimental section. Fig. 6 schematically indicates the places in the glovebox where the indicators were located (positions 1 to 10). All the chemical indicators proved bio-decontamination, by changing their colour from purple to rose during the bio-decontamination cycle. After 7 days, the culture medium remained purple in all tubes containing the biological indicators exposed to the bio-decontamination cycle indicating a 6-Log reduction (that is, a 99.9999% reduction) in the Geobacillus stearothermophilus spore concentration. Only the nonexposed biological indicator (control tube) was yellow after 24 h, indicating microbial growth. To obtain a more quantitative characterization of the bio-decontamination efficacy of H₂O₂ loaded in SG60, D-values were determined. The D-value is the time required to achieve a log reduction, that is, to kill 90% of the present microorganisms. Three sets of experiments with different H₂O₂ vapour profiles were carried out (Fig. 7). Biological indicators containing 1 million spores were exposed to H₂O₂ vapour and the number of surviving spores was measured by the Most Probable Number Method, as described in the experimental section. After an initial 5-min lag phase, in which little sporicidal activity was detected, a rapid phase of spore killing ensued (Fig. 7b). This lag phase was probably caused by the time needed to build up H₂O₂ levels in the vapour phase. In the fast phase, D-values of 1 2 can be calculated depending on the precise conditions, which compares very well with H₂O₂ generated from liquid solutions²⁹.Thus, after being released into the vapour phase, the H₂O₂ loaded in SG60 has a very high bio-decontamination efficacy.


3.5. Transportation

Transportation safety is an important issue when dealing with concentrated solutions of H₂O₂. As liquid H₂O₂ is classified as an Oxidizing Substance (Division 5.1, Oxidizing Liquid)³⁰, belonging to the Packing group II, the H₂O₂ loaded SG60 was tested. Relevant samples were subjected to Division 5.1 Oxidizer Analysis in accordance with the United Nations Transport of Dangerous Goods- Manual of Tests and Criteria, seventh revised edition (2019)³¹. The tests were performed by an external ISO 2001:2015 Certified Laboratory, which concluded that the material does not appear to be a Division 5.1 Solid Oxidizer, as defined by the United Nations and The United States Department of Transportation criteria. Additional information on the methodology and test results can be found in Supplementary Information (Section 7).

Fig. 6. Locations of the chemical and biological (Geobacillus stearothermophilus spore discs) indicators in the glovebox.
Fig. 6. Locations of the chemical and biological (Geobacillus stearothermophilus spore discs) indicators in the glovebox.
Fig. 7. Determination of D-values: a) profile of H₂O₂ vapour; b) profile of log10 of surviving Geobacillus stearothermophilus spores. The lines correspond to three conditions labelled according to the peak H₂O₂ concentration reached: 379 ppm, 468 ppm and 2578 ppm.
Fig. 7. Determination of D-values: a) profile of H₂O₂ vapour; b) profile of log10 of surviving Geobacillus stearothermophilus spores. The lines correspond to three conditions labelled according to the peak H₂O₂ concentration reached: 379 ppm, 468 ppm and 2578 ppm.

3.6. Reusability of SG60

Today, circular economy is key issue for economic and environmental sustainability. Thus, SG60 material utilized in biodecontamination cycles was subjected to hydrogen peroxide recharging to assess its viability for reuse. Table 3 provides data on H₂O₂ levels in both new SG60 and reused SG60, silica that has been recycled after the loading process. Each row in the table corresponds to a separate batch.

The evidence suggests that SG60 effectively retains its loading capacity after utilization, thereby enabling a sustained and repeated recycling process. Thus, the material can be employed multiple times without a significant decline in its capacity to retain H₂O₂, offering practical advantages in terms of resource utilization and cost effectiveness.


Table 3

H₂O₂ content in new and reused SG60 material after loading (continuous mode).

3.7. Bio-decontamination of biological molecules

Biopharmaceuticals are an emerging class of drugs in which microbial decontamination poses a problem. It is usually assumed that biodecontamination protocols inactivate these products. Our proof of concept with the enzyme deoxyribonuclease (DNAse) experiment shows that the bio decontamination of these products is possible without loss of activity.

DNAse is a type of nuclease enzyme that catalyses the hydrolytic cleavage of phosphodiester linkages in the DNA molecule In Fig. 8, the activity of DNAase is followed as variation in absorbance at 260nm (ΔAbs 260nm) along time for the test samples and controls. As expected, for the heat inactivated protein sample, DNA cleavage did not occur and thus no significant variation in absorbance was verified. For the enzyme samples exposed to 100 mg of H₂O₂ loaded material, a variation in absorbance was verified, indicating that the enzyme maintained its activity. Log 6 bio-decontamination was achieved in 2 out of the 3 samples, shown by the negative biological indicators after incubation. For the enzyme samples exposed to 300 mg of H₂O₂, while H₂O₂ inactivated the biological indicators, a loss of DNase activity was observed in all samples.

Although the mass of H₂O₂-loaded material should be optimized to ensure the 6-log the bio decontamination without compromising the enzyme integrity, these experiments show the potential for the low temperature bio-decontamination.



Conclusion


In this work we showed for the first time that H₂O₂ can be stored upon adsorption from vapour phase in a nanoporous material and later, after gentle heating at temperatures between 60 and 80°C, rapidly released in the gas phase in concentrations that enable efficient biodecontamination.

Various types of nanoporous materials can be used for this end, from zeolites to mesoporous silicas, including relatively cheap materials such as Silica Gel 60 (SG60). The stability of the H₂O₂ loaded in SG60 is modest at room temperature but can be preserved in the freezer (at -18°C) for over six months. For samples where the SG60 is mixed with EDTA or citric acid in concentrations between 1 and 2%, the stability of the adsorbed H₂O₂ is clearly improved. We proved that the silica SG60 can be reused after loading and release of H₂O₂, maintaining its capacity of H₂O₂ retention. In addition, we also showed that at subzero temperatures there is a slow release of H₂O₂ vapour from the porous material that can be used to decontaminate biological products without loss of activity. In conclusion, the utilization of nanoporous materials for the adsorption and subsequent desorption of H₂O₂ represents an innovative and simple method for an efficient use of H₂O₂ in its vapour phase, enhancing current bio decontamination procedures and opening the field to new possible applications, such as the bio decontamination of biopharmaceuticals.


Fig. 8. Bio-decontamination of DNAse containing vials without loss of enzymatic activity Tubes (50 mL) containing DNAse were exposed to 100 mg and 300 mg of SG60 loaded with H₂O₂ for 1 week in a freezer (-18°C). Both amounts were sufficient to achieve a 6-log decontamination. Samples with untreated DNAse (DNAse) and heat Inactivated DNAse (70°C for 15 min) are also shown.
Fig. 8. Bio-decontamination of DNAse containing vials without loss of enzymatic activity Tubes (50 mL) containing DNAse were exposed to 100 mg and 300 mg of SG60 loaded with H₂O₂ for 1 week in a freezer (-18°C). Both amounts were sufficient to achieve a 6-log decontamination. Samples with untreated DNAse (DNAse) and heat Inactivated DNAse (70°C for 15 min) are also shown.

Credit authorship contribution statement

Fadhil Musa: Writing – review & editing, Validation, Project administration, Investigation, Funding acquisition, Formal analysis.

Raquel Nogueira: Methodology, Investigation. Margarida Beiral: Writing – review & editing, Methodology, Investigation, Formal analysis. Fernando Antunes: Writing – review & editing, Writing – original draft, Validation, Supervision, Resources, Methodology, Investigation, Funding acquisition, Formal analysis, Data curation, Conceptualization. João Pires: Writing – original draft, Methodology, Investigation, Formal analysis, Conceptualization.


Declaration of competing interest

The authors declare the following financial interests/personal relationships which may be considered as potential competing interests: JP, FA, and FM are founders of Delox Lda., a company that was created to develop applications in bio-decontamination, using H₂O₂ desorbed from nanoporous materials If there are other authors, they declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.


Delox has developed a novel, patented sterilization agent based on a solid formulation that releases Hydrogen Peroxide Vapor and a device that will bio-decontaminate volumes up to 2 m³ (70 ft³). Delox technology concentrates and releases hydrogen peroxide vapor at low temperatures (60°C), which simplifies the sterilization device design, reducing power consumption and equipment size and cost, compared to other solutions. Furthermore, Delox has developed a highly efficient aeration module that enables a rapid decomposition process and air circulation system to distribute the hydrogen peroxide throughout the enclosure. Delox has refined its hardware and methods to enable a complete 6-log bio-decontamination laboratory equipment, isolators and biosafety cabinets. www.delox-global.com


This article, republished from Microporous and Mesoporous Materials under the terms of the Creative Commons Attribution 4.0 International License (CC BY 4.0), can be accessed via DOI: 10.1016/j.micromeso.2024.110341 . The original publication is credited, and no restrictions beyond those permitted by the license apply. Download original article published in Microporous and Mesoporous Materials 


Data availability

Data will be made available on request.


Acknowledgements

Financed by Fundação para a Ciências e a Tecnologia, I.P./MCTES through national funds (PIDDAC) CQE - UIDB/00100/2020 and UIDP/ 00100/2020 and IMS - LA/P/0056/2020.


Appendix A. Supplementary data

Supplementary data to this article can be found online at https://doi.org/10.1016/j.micromeso.2024.113056. This article was originally published in Microporous and Mesoporous Materials in 2024.



References

[1] G. Kampf, D. Todt, S. Pfaender, E. Steinmann, Persistence of coronaviruses on

inanimate surfaces and their inactivation with biocidal agents, J. Hosp. Infect. 104

(2020) 246–251, https://doi.org/10.1016/j.jhin.2020.01.022.

[2] B. McEvoy, N.J. Rowan, Terminal sterilization of medical devices using vaporized

hydrogen peroxide: a review of current methods and emerging opportunities,

J. Appl. Microbiol. 127 (2019) 1403–1420, https://doi.org/10.1111/jam.14412.

[3] J.M. Boyce, Modern technologies for improving cleaning and disinfection of

environmental surfaces in hospitals, Antimicrob. Resist. Infect. Control 5 (2016),

[4] S.J. Dancer, Controlling hospital-acquired infection: focus on the role of the

environment and new technologies for decontamination, Clin. Microbiol. Rev. 27

(2014) 665–690, https://doi.org/10.1128/CMR.00020-14.

[5] E. Eschlbeck, C. Seeburger, U. Kulozik, Spore inactivation on solid surfaces by

vaporized hydrogen peroxide-Influence of carrier material surface properties,

J. Food Sci. 85 (2020) 1536–1541, https://doi.org/10.1111/1750-3841.15086.

[6] M.A. Czarneski, K. Lorcheim, A discussion of biological safety cabinet

decontamination methods: formaldehyde, chlorine dioxide, and vapor phase

hydrogen peroxide, Appl. Biosaf. 16 (2011) 26–33, https://doi.org/10.1177/

153567601101600104.

[7] J. Spiegelman, D. Alvarez, Cheating Raoult’s law to enable delivery of hydrogen

peroxide as a stable vapor, Gases Instrum. January/Fe (2015) 14–19.

[8] M.H. Sk, R.A. Overfelt, R.L. Haney, J.W. Fergus, Hydrogen embrittlement of 4340

steel due to condensation during vaporized hydrogen peroxide treatment, Mater.

Sci. Eng. A-STRUCTURAL Mater. Prop. Microstruct. Process. 528 (2011)

3639–3645, https://doi.org/10.1016/j.msea.2011.01.100.

[9] J.A. Kumar, J.L. Cadnum, A.L. Jencson, C.J. Donskey, Efficacy of a multi-purpose

high level disinfection cabinet against Candida auris and other health careassociated

pathogen, Am. J. Infect. Control 48 (2020) 849–850, https://doi.org/

10.1016/j.ajic.2019.11.029.

[10] M. Thommes, K. Kaneko, A. V Neimark, James P. Olivier, F. Rodriguez-Reinoso,

J. Rouquerol, K.S.W. Sing, Physisorption of gases, with special reference to the

evaluation of surface area and pore size distribution (IUPAC Technical Report),

Pure Appl. Chem. 87 (2015) 1051–1069, https://doi.org/10.1515/pac-2014-1117.

[11] F. Bi, S. Ma, B. Gao, B. Liu, Y. Huang, R. Qiao, X. Zhang, Boosting toluene deep

oxidation by tuning metal-support interaction in MOF-derived Pd@ZrO2 catalysts:

the role of interfacial interaction between Pd and ZrO2, Fuel 357 (2024) 129833,

[12] M. Thommes, C. Schlumberger, Characterization of nanoporous materials, Annu.

Rev. Chem. Biomol. Eng. 12 (2021) 137–162, https://doi.org/10.1146/annurevchembioeng-

061720-081242.

[13] V.K. Saini, M. Andrade, M.L. Pinto, A.P. Carvalho, J. Pires, How the adsorption

properties get changed when going from SBA-15 to its CMK-3 carbon replica, Sep.

Purif. Technol. 75 (2010) 366–376, https://doi.org/10.1016/j.

seppur.2010.09.006.

[14] T. Borrego, M. Andrade, M.L. Pinto, A. Rosa Silva, A.P. Carvalho, J. Rocha,

C. Freire, J. Pires, Physicochemical characterization of silylated functionalized

materials, J. Colloid Interface Sci. 344 (2010) 603–610, https://doi.org/10.1016/j.

jcis.2010.01.026.

[15] B. Unger-Bimczok, V. Kottke, C. Hertel, J. Rauschnabel, The influence of humidity,

hydrogen peroxide concentration, and condensation on the inactivation of

Geobacillus stearothermophilus spores with hydrogen peroxide vapor, J. Pharm.

Innov. 3 (2008) 123–133, https://doi.org/10.1007/s12247-008-9027-1.

[16] J.R. Haines, B.A. Wrenn, E.L. Holder, K.L. Strohmeier, R.T. Herrington, A.

D. Venosa, Measurement of hydrocarbon-degrading microbial populations by a 96-

well plate most-probable-number procedure, J. Ind. Microbiol. Biotechnol. 16

(1996) 36–41, https://doi.org/10.1007/BF01569919.

[17] M.H. McCrady, The numerical interpretation of fermentation-tube results, J. Infect.

Dis. 17 (1915) 183–212. http://www.jstor.org/stable/30083495.

[18] D. Lewandowski, D. Bajerlein, G. Schroeder, Adsorption of hydrogen peroxide on

functionalized mesoporous silica surfaces, Struct. Chem. 25 (2014) 1505–1512,

[19] P.A. Wright, G.M. Pearce, Structural chemistry of zeolites, in: J. Cejka, A. Corma,

S. Zones (Eds.), Zeolites Catal. Synth. React. Appl., WILEY-VCH Verlag GmbH &

Co., Weinheim, 2010, pp. 171–208.

[20] W. Lukens Wayne, P. Schmidt-Winkel, D. Zhao, J. Feng, G.D. Stucky, Evaluating

pore sizes in mesoporous materials: a simplified standard adsorption method and a

simplified Broekhoff􀀀 de Boer method, Langmuir 15 (1999) 5403–5409, https://

[21] A. Petushkov, N. Ndiege, A.K. Salem, S.C. Larsen, in: J.C. Fishbein (Ed.), Chapter 7

- Toxicity of Silica Nanomaterials: Zeolites, Mesoporous Silica, and Amorphous

Silica Nanoparticles, Elsevier, 2010, pp. 223–266, https://doi.org/10.1016/S1872-

0854(10)04007-5.

[22] A. Plauck, E.E. Stangland, J.A. Dumesic, M. Mavrikakis, Active sites and

mechanisms for H2O2 decomposition over Pd catalysts, Proc. Natl. Acad. Sci. USA

113 (2016) E1973–E1982, https://doi.org/10.1073/pnas.1602172113.

[23] S. Croft, B.C. Gilbert, J.R.L. Smith, J.K. Stell, W.R. Sanderson, Mechanisms of

peroxide stabilization. An investigation of some reactions of hydrogen peroxide in

the presence of aminophosphonic acids, J. Chem. Soc.{,} Perkin Trans. 2 (1992)

153–160, https://doi.org/10.1039/P29920000153.

[24] F. Kelly, C. McKay, H. Steed, Brian, Solutions for Stabilizing Hydrogen Peroxide

Containing Solutions, 1991. WO/1991/008981.

[25] A.G. Leigh, Stabilization of Hydrogen Peroxide, 1981 US4304762.

[26] E.F. Dougherty, Q.G. Hopkins, J.E. Sanborn, A.T. Weibel, Tin (II) Stabilizer for

Hydrogen Peroxide, 1992. WO/1992/003377.

[27] T. Boonpoempoon, W. Wonsawat, T. Kaneta, Long-term stabilization of hydrogen

peroxide by poly(vinyl alcohol) on paper-based analytical devices, Sci. Rep. 9

(2019) 12951, https://doi.org/10.1038/s41598-019-49393-6.

[28] U.S. Environmental Protection Agency, Low-oncentration Hydrogen Peroxide

(LCHP) Vapor for Bioremediation, 2017.

[29] M. Sawale, X. Cheng, R. Drolia, P. Benyathiar, F. Ozadali, A. Bhunia, D. Mishra,

Inactivation kinetics of Bacillus atrophaeus in liquid hydrogen peroxide for aseptic

package sterilization, Lebensm. Wiss. Technol. 170 (2022) 114074, https://doi.

org/10.1016/j.lwt.2022.114074.

[30] Class 5.1, [Oxidizing substances]: hydrogen peroxide, European Council of

Chemical Manufacturers’ Federations, 1989.

[31] United Nations, Recommendations on the Transport of Dangerous Goods Manual of

Tests and Criteria, seventh ed., 2019, pp. 382–385.


Authors

Fadhil Musa¹, Raquel Nogueira¹, Margarida Beiral¹, Fernando Antunes², João Pires²*


¹Delox Lda, Tec Labs, Campus da FCUL, Campo Grande, 17649-016, Lisboa, Portugal

²CQE, Centro de Química Estrutural, Institute of Molecular Sciences, Departamento de Química e Bioquímica, Faculdade de Ciências, Universidade de Lisboa, Campo Grande, 1749-016, Lisboa, Portugal


* Corresponding author:

(J. Pires). jpsilva@fc.ul.pt




 
 
 

Comments


bottom of page